I-STAT CHEM8+ CARTRIDGE
Received May 27, 2026 · Event occurred May 18, 2026
Report 2245578-2026-00119 · MDR key 25301643
Device
Generic name
Chem8+ Cartridge
Manufacturer
Abbott Point Of CareModel number
NACatalog number
09P31-26
Lot number
H26085
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
30 YR · Female
- Limb Pain
Narrative
Description of Event or Problem
ON 19-MAY-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT CREATININE RESULT ON A 30-YEAR-OLD FEMALE PATIENT WITH LEFT ARM PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD / DATE / TESTED / RESULTS / SAMPLE: I-STAT, (B)(6) 2026, 17:48, 3.7 MG/DL, A; VITROS 7600, (B)(6) 2026, 19:05, 0.6 MG/DL, B. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.
Additional Manufacturer Narrative
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.