ORCA?
Received May 26, 2026 · Event occurred Mar 9, 2026
Report 3005099803-2026-02084 · MDR key 25285392
Device
Generic name
Endoscope Channel Accessory
Manufacturer
Boston Scientific CorporationModel number
SUV-617Catalog number
SUV-617
Lot number
0038400303
Product problems
- Infusion or Flow Problem
- Physical Resistance/Sticking
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
BLOCK H6: IMDRF A0509 CAPTURES THE REPORTABLE EVENT OF SUCTION VALVE STICKING. BLOCK H7: ON MAY 12, 2026, BOSTON SCIENTIFIC INITIATED A FIELD SAFETY CORRECTIVE ACTION (FSCA - BOSTON SCIENTIFIC REFERENCE: (B)(4)) THAT INCLUDED THE REMOVAL OF ORCA? SINGLE USE AIR/WATER AND SUCTION VALVES ASSOCIATED WITH BATCH NUMBER 38400303 DUE TO INCREASED REPORTS OF SUCTION BUTTON STICKING ISSUES. A COMMUNICATION WAS SENT OUT TO MATERIALS MANAGERS/HEALTH CARE PROFESSIONALS ON (B)(6) 2026, WITH INSTRUCTIONS TO IMMEDIATELY STOP FURTHER USE OR DISTRIBUTION OF ORCA? SINGLE USE AIR/WATER AND SUCTION VALVES ASSOCIATED WITH BATCH NUMBER 38400303, REMOVE THE DEVICES FROM INVENTORY, AND SEGREGATE THEM IN A SECURE LOCATION UNTIL THEY CAN BE RETURNED TO BOSTON SCIENTIFIC. BOSTON SCIENTIFIC WILL CONTINUE TO CLOSELY MONITOR AND TREND SIMILAR EVENTS AND ASSOCIATED RISKS, AS OUTLINED IN OUR QUALITY SYSTEMS PROCESS.
Description of Event or Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT ORCA AIR WATER AND SUCTION VALVES WERE USED IN A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2026. THE SUCTION VALVE BECAME STUCK, AND THE BUTTONS HAD TO BE REPLACED WITH METAL BUTTONS. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Remedial action
- Recall