inforMED
MalfunctionODC

ORCA?

Received May 26, 2026 · Event occurred Mar 12, 2026

Report 3005099803-2026-02078 · MDR key 25280790

Device

Generic name

Endoscope Channel Accessory

Model number

SUV-617-50

Catalog number

SUV-617-50

Product problems

  • Physical Resistance/Sticking

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BLOCK D4 AND H4: THE COMPLAINANT WAS UNABLE TO PROVIDE THE SUSPECT DEVICE LOT NUMBER; THEREFORE, THE LOT EXPIRATION AND DEVICE MANUFACTURE DATES ARE UNKNOWN. BLOCK H6: IMDRF DEVICE CODE A0509 CAPTURES THE REPORTABLE EVENT OF SUCTION VALVE STICKING. BLOCK H7, H9, H11: ON MAY 12, 2026, BOSTON SCIENTIFIC INITIATED A FIELD SAFETY CORRECTIVE ACTION (FSCA - BOSTON SCIENTIFIC REFERENCE: 97585870-FA) THAT INCLUDED THE REMOVAL OF ORCA? SINGLE USE AIR/WATER AND SUCTION VALVES ASSOCIATED WITH BATCH NUMBER 38400303 DUE TO INCREASED REPORTS OF SUCTION BUTTON STICKING ISSUES. A COMMUNICATION WAS SENT OUT TO MATERIALS MANAGERS/HEALTH CARE PROFESSIONALS ON MAY 12, 2026, WITH INSTRUCTIONS TO IMMEDIATELY STOP FURTHER USE OR DISTRIBUTION OF ORCA? SINGLE USE AIR/WATER AND SUCTION VALVES ASSOCIATED WITH BATCH NUMBER 38400303, REMOVE THE DEVICES FROM INVENTORY, AND SEGREGATE THEM IN A SECURE LOCATION UNTIL THEY CAN BE RETURNED TO BOSTON SCIENTIFIC. BOSTON SCIENTIFIC WILL CONTINUE TO CLOSELY MONITOR AND TREND SIMILAR EVENTS AND ASSOCIATED RISKS, AS OUTLINED IN OUR QUALITY SYSTEMS PROCESS.

Description of Event or Problem

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORCA VALVE WAS USED DURING PROCEDURE PERFORMED ON (B)(6) 2026. THE SUCTION BUTTONS WERE STICKING BOTH BEFORE AND DURING THE PROCEDURE. EVEN AFTER SOAKING THE BUTTONS IN WATER, APPLYING LUBRICANT, AND REDUCING THE SUCTION LEVEL, THE ISSUE CONTINUED TO OCCUR INTERMITTENTLY. SOME BUTTONS BECAME STUCK ON THE FIRST PRESS, WHILE OTHERS FUNCTIONED NORMALLY THROUGHOUT THE PROCEDURE. THE PROCEDURE WAS ULTIMATELY COMPLETED USING A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Remedial action

  • Recall