inforMED
InjuryHST

TOWER PARTS ASSEMBLY

Received May 26, 2026 · Event occurred Apr 20, 2026

Report 3025141-2026-00267 · MDR key 25280260

Device

Generic name

Apparatus, Traction, Non-powered

Manufacturer

Acumed Llc

Model number

10100001

Lot number

663594

Product problems

  • Mechanical Jam
  • Damaged Thread

Patient

34 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED: SCREW CROSS-THREADED ON TRACTION TOWER. TOWER WAS ASSEMBLED BY SURGEON. "KNOB WOULD NOT TURN" PER SURGEON. PROCEDURE WAS COMPLETED WITH MANUAL GROSS TRACTION. CAUSED 15 MINUTE DELAY.

Additional Manufacturer Narrative

EXAMINATION CONFIRMED REPORTED CROSS-THREADING OF THE RETURNED COMPONENTS OF THE TRACTION TOWER. HOWEVER NO DEFINITIVE CONCLUSION COULD BE MADE AS TO THE CAUSE OF THE CROSS-THREADING.