VIDEO-OPTIK "ENDOEYE 3D", 30°

Received May 22, 2026

Report 9610773-2026-03976 · MDR key 25262752

Device

Generic name

Rigid Video Laparoscope

Model number

WA50082A

Product problems

  • Poor Quality Image

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT DURING INSPECTION FOR USE, THE RIGID VIDEO LAPAROSCOPE EXHIBITED POOR IMAGE TRANSMISSION. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem

NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

Additional Manufacturer Narrative

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELD: D8, D9, H2, H3, H4, H6, H11 THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS NOT CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: THE LASER LIGHT CABLE PLUG OF THE VIDEO CABLE SECTION CONTAINS FOREIGN MATERIAL. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE CAUSE OF THE CUSTOMER¿S ALLEGATION COULD NOT BE DETERMINED. IN ADDITION, FOR THE ADDITIONAL FINDING, THE CAUSE COULD NOT BE ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.