inforMED
MalfunctionQCA

NEUROFORM ATLAS 3.0MM X 15MM NO TIP - CE

Received May 22, 2026 · Event occurred May 11, 2026

Report 3008881809-2026-00330 · MDR key 25257094

Device

Generic name

Intracranial Coil-assist Stent

Catalog number

M003EZAS30150

Lot number

25960849

Product problems

  • Premature Activation

Patient

59 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE STENT (SUBJECT DEVICE) WAS USED IN THE MEDICAL PROCEDURE. HOWEVER, WHEN HALF OF THE STENT ENTERED THE MICROCATHETER FURTHER DELIVERY BECAME DIFFICULT. THE PHYSICIAN ADJUSTED THE DELIVERY WIRE AND FOUND THE SUBJECT STENT TO BE DEPLOYED WITHIN THE PROXIMAL END OF MICROCATHETER. THE SUBJECT DEVICE WAS REPLACED ALONG WITH THE MICROCATHETER, AND THE PROCEDURE WAS REPORTED TO BE COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT.