MalfunctionDRG

ORG-9110A

Received May 21, 2026 · Event occurred May 7, 2026

Report 8030229-2026-06606 · MDR key 25249775

Device

Generic name

Multiple Patient Receiver

Model number

ORG-9110A

Catalog number

ORG-9110A

Lot number

NA

Product problems

  • Communication or Transmission Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE MULTIPLE PATIENT RECEIVER (ORG) HAD A LIT ERROR LIGHT AND DROPPED THE SIGNAL TO THE ZM TRANSMITTERS. NO PATIENT HARM WAS REPORTED.

Additional Manufacturer Narrative

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE COMMUNICATION LOSS IS AFFECTING EIGHT TELEMETRY UNITS ON THEIR CNS-6201. HE WOULD LIKE TO SEND THE MULTIPLE PATIENT RECEIVER (ORG) IN FOR REPAIR BECAUSE THE ERROR LIGHT REMAINS ILLUMINATED ON THE UNIT. HE ATTEMPTED TO POWER CYCLE THE ORG, BUT THE ERROR LIGHT DID NOT CLEAR. BIOMED USED A SPARE ORG TO RESTORE MONITORING FOR THE EIGHT TELEMETRY UNITS THAT HAD GONE INTO COMMUNICATION LOSS. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.