VIDEO-OPTIK "ENDOEYE 3D", 30°
Received May 21, 2026 · Event occurred May 5, 2026
Report 9610773-2026-03947 · MDR key 25247909
Device
Product problems
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE MOST PROBABLE CAUSE OF THE MALFUNCTION IS TRACED TO COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE EXHIBITED BROKEN CHARGED COUPLED DEVICE UNIT. THERE WAS NO PATIENT INVOLVEMENT.