Received May 21, 2026 · Event occurred May 13, 2026
Report 3042177665-2026-000083 · MDR key 25239280
Device
Generic name
Full-body Mri System, Superconducting Magnet
Manufacturer
Philips Medical Systems Nederland B.v.Model number
SMARTPATH TO DSTREAM FOR XR AND 3.0TCatalog number
782113
Product problems
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
PHILIPS RECEIVED A REPORT THAT THE QBC SEAL HAS COME LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL COMES LOOSE, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.
Additional Manufacturer Narrative
THE QBC SEAL HAS COME LOOSE, WAS REPLACED AND THE SYSTEM WAS HANDED BACK TO THE CUSTOMER FOR USE. THE FAILURE OF THE QBC SEAL TO REMAIN IN PLACE IS CONSIDERED A MALFUNCTION. REMEDIAL ACTION: A FIELD ACTION WAS INITIATED UNDER 2023-PD-MR-013. A FIELD SAFETY NOTIFICATION WAS SENT TO THE CUSTOMERS INFORMING THEM ABOUT THIS POTENTIAL ISSUE. THE FIELD CORRECTION HAS SINCE BEEN MADE AVAILABLE TO THE FIELD IN DEC 2024.
Remedial action
- Modification/Adjustment