inforMED
MalfunctionJGS

I-STAT CHEM8+ CARTRIDGE

Received May 20, 2026 · Event occurred May 11, 2026

Report 2245578-2026-00114 · MDR key 25226212

Device

Generic name

Chem8+ Cartridge

Model number

NA

Catalog number

09P31-25

Lot number

H25318A

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

APOC INCIDENT # (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem

ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT POTASSIUM RESULT ON A PATIENT. THERE WAS NO PATIENT INFORMATION, NOR DETAILS SURROUNDING THE EVENT AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: RESULT : SAMPLE: I-STAT, 6.0 MMOL/L, WHOLE BLOOD, LABORATORY, 3.6 MMOL/L , SERUM. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.