inforMED
InjuryNHW

PATCH

Received Apr 9, 2012 · Event occurred Feb 23, 2012

Report 2124215-2012-03040 · MDR key 2522310

Add to watch list
Product code (NHW)
Model (0041)

Device

Generic name

Implantable Lead

Model number

0041

Product problems

  • High impedance
  • Device Operates Differently Than Expected

Patient

43 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THESE TWO EPICARDIAL PATCHES WERE EXHIBITING HIGH SHOCKING IMPEDANCE MEASUREMENTS GREATER THAN 200 OHMS. IT WAS UNKNOWN WHICH PATCH WAS RESPONSIBLE. HOWEVER, IT WAS REPORTED THAT THE REAR PATCH WAS BEGINNING TO SEPARATE FROM THE HEART. A REVISION PROCEDURE WAS PERFORMED AND BOTH PATCHES WERE SURGICALLY ABANDONED. A NEW SYSTEM WAS IMPLANTED SUB MAMMARY. NO INVESTIGATION PLANNED TO BE PERFORMED TO DETERMINE WHICH EPICARDIAL PATCH WAS RESPONSIBLE FOR THE HIGH IMPEDANCE MEASUREMENTS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Additional Manufacturer Narrative

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.