VIDEO-OPTIK "ENDOEYE 3D", 30°

Received May 19, 2026 · Event occurred May 7, 2026

Report 9610773-2026-03842 · MDR key 25216834

Device

Generic name

Rigid Video Laparoscope

Model number

WA50082A

Product problems

  • Loose or Intermittent Connection
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: THE RESOLUTION IS INADEQUATE. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT THE REPORTED EVENT OCCURRED DUE TO DAMAGE TO THE LLK (LASER LIGHT CABLE) PLUG OF THE VIDEO CABLE SECTION. ADDITIONALLY, THE INADEQUATE RESOLUTION IS LIKELY DUE TO A BROKEN R-UNIT (CHARGED COUPLED DEVICE). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

IT WAS REPORTED THAT THE LOCKING SPRING OF THE RIGID VIDEO LAPAROSCOPE PLUG WAS MISSING, CAUSING THE PLUG TO BECOME LOOSE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.