inforMED
MalfunctionILZ

PATIENT HELPER ADAPTER (HEIGHT ADJ)

Received Apr 4, 2012 · Event occurred Mar 1, 2012

Report 1035617-2012-00002 · MDR key 2521021

Device

Generic name

Patient Helper Adapter (height Adj)

Manufacturer

Zimmer Surgical

Model number

NA

Catalog number

00-0965-002-01

Lot number

NA

Product problems

  • Bent

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE SLEEVE SECTION OF THE PT HELPER BENT WHERE IT ATTACHES TO THE BED BRACKET. ADDITIONAL CLINICAL FOLLOW UP WITH THE HOSPITAL INDICATED THAT THE PT HELPER HAD BEEN ON THE PT'S BED PROVIDING TRAPEZE USE FOR QUITE SOME TIME. THE RN STATED THE PT WAS CONSIDERED TO BE QUITE PROACTIVE WITH PHYSICAL THERAPY AND SELF EXERCISE TO REGAIN LIMB FUNCTION AND STRENGTH. THE RN STATED THE PT WAS "DOING CHIN-UPS ON THE TRAPEZE, LIFTING HIS FULL BODY WEIGHT COMPLETELY OFF THE BED, WHEN THE PT NOTED THE BAR TO BE BENDING SLIGHTLY OVER THE BED." THE RN STATED THE PT WAS NOT HARMED AND NEITHER THE TRAPEZE NO THE PT HELPER FELL.

Additional Manufacturer Narrative

THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER AT THE TIME OF THIS REPORT. A FOLLOW UP MEDWATCH WILL BE SUBMITTED ONCE THE DEVICE HAS BEEN RETURNED AND THE INVESTIGATION IS COMPLETE.