inforMED
MalfunctionFLE

AMSCO

Received May 15, 2026 · Event occurred Apr 17, 2026

Report 3005899764-2026-00037 · MDR key 25188546

Device

Generic name

Sterilizer

Catalog number

SR0201110311

Product problems

  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED WATER AND STEAM LEAKING FROM THEIR AMSCO 400 STERILIZER ONTO THE FLOOR. NO REPORT OF INJURY.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE UNIT AND FOUND THE VIEWPORT WAS DAMAGED ALLOWING WATER TO LEAK FROM THE STERILIZER. THE TECHNICIAN REPLACED THE VIEWPORT, TESTED THE UNIT, CONFIRMED THE UNIT TO BE OPERATING ACCORDING TO SPECIFICATION, AND THE UNIT WAS RETURNED TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.