inforMED
MalfunctionNIG

ALINITY I CA 19-9XR REAGENT KIT

Received May 14, 2026 · Event occurred Mar 24, 2026

Report 3002809144-2026-00122 · MDR key 25176594

Device

Generic name

System, Test, Carbohydrate Ag (ca19-9), For Monit And Mgmt Of Pancreatic Cancer

Manufacturer

Abbott Gmbh

Catalog number

08P32-20

Lot number

83893FP00

Product problems

  • Imprecise or Erratic Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER OBSERVED IMPRECISE RESULTS FOR THE ALINITY CA 19-9XR FOR 3 PATIENTS. CLINICAL HISTORY IS UNKNOWN FOR THE PATIENTS. THE FOLLOWING DATA WAS PROVIDED: 73-YEAR-OLD MALE MULTIPLE SAMPLES. CUSTOMER IS QUESTIONING THE RESULT FROM (B)(6) 2026: SID (B)(6) PROCESSED ON (B)(6) 2026 = 1034.27 U/ML SID (B)(6) PROCESSED ON (B)(6) 2026 = 270.59 U/ML SID (B)(6) PROCESSED ON (B)(6) 2026 = 236.94 U/ML HISTORICAL RESULTS: SID (B)(6) PROCESSED ON (B)(6) 2026 = 169.39 U/ML SID (B)(6) PROCESSED ON (B)(6) 2026 = 180.56 U/ML SID (B)(6) PROCESSED ON (B)(6) 2026 = 160.12 U/ML 25-YEAR-OLD FEMALE SID (B)(6) INITIAL RESULT = 40.82 REPEAT RESULT = 74.56 U/ML 63-YEAR-OLD FEMALE SID (B)(6) INITIAL RESULT = 57.31 REPEAT RESULT = 85.83 U/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS PROVIDED, NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE.