inforMED
MalfunctionGES

BARD-PARKER DISPOSABLE SCALPELS SIZE 11, NON-STERILE

Received May 14, 2026 · Event occurred Mar 26, 2026

Report 3002750084-2026-00046 · MDR key 25174175

Device

Generic name

Scalpel Blade

Model number

371631

Lot number

UNKNOWN

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. A PICTURE WAS PROVIDED, BUT THE COMPLAINT COULD NOT BE CONFIRMED BASED ON WHAT WAS IN THE PICTURE. THE MOST LIKELY ROOT CAUSE IS A MANUFACTURING ERROR, HOWEVER WITHOUT A SAMPLE TO EVALUATE IT CANNOT BE CONFIRMED. USER ERROR MAY ALSO HAVE CAUSED THE BLADE BREAKAGE DUE TO EXCESSIVE FORCE. THIS SHOULD BE CONSIDERED THE FINAL REPORT. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem

COMPLAINANT ALLEGES "SCALPEL BLADE BROKE OFF IN PATIENT, MOUNT FOR BLADE TO HANDLE BROKE. THE SCALPEL BROKE OFF IN THE PATIENT WHILE CUTTING TISSUES AND HAD TO BE REMOVED."