inforMED
MalfunctionFMT

F&P NEOPUFF

Received May 14, 2026

Report 9611451-2026-01610 · MDR key 25167902

Add to watch list
Product code (FMT)
Manufacturer (Fisher & Paykel Healthcare, Ltd.)
Model (RD900AEU)

Device

Generic name

Infant Resuscitator

Model number

RD900AEU

Catalog number

RD900AEU

Lot number

002691

Product problems

  • Loose or Intermittent Connection
  • Pressure Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

(B)(4). SECTION D4: UDI DETAILS ARE NOT AVAILABLE BASED ON THE TIME OF MANUFACTURING. METHOD: THE SUBJECT DEVICE, RD900AEU NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTER IN CALIFORNIA WHERE IT WAS INSPECTED BY A TRAINED F&P SERVICE TECHNICIAN. OUR INVESTIGATION IS THUS BASED ON THE INFORMATION PROVIDED BY THE F&P SERVICE TECHNICIAN, THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: UPON EVALUATION OF THE SERVICE REPORT, IT WAS CONFIRMED THAT THE MANOMETER HAD A LOOSE SCREW AND IT FAILED CALIBRATION TESTING. CONCLUSION: WE ARE UNABLE TO DETERMINE THE EXACT CAUSE OF THE REPORTED EVENT. AS PART OF MANUFACTURING PROCESS, EACH NEOPUFF UNIT UNDERGOES 100% TESTING ON THE PRODUCTION LINE TO ENSURE COMPLIANCE WITH CRITICAL PRODUCT SPECIFICATIONS. UNITS THAT DO NOT MEET THE SPECIFICATIONS ARE REJECTED. THE SUBJECT NEOPUFF WOULD HAVE MET THE REQUIREMENTS AT THE TIME OF PRODUCTION. THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. IN ADDITION, THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF". AS MENTIONED, THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING:- DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT.

Description of Event or Problem

DURING A DEVICE EVALUATION AT THE FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTER IN CALIFORNIA, THE MANOMETER OF AN RD900AEU NEOPUFF INFANT RESUSCITATOR FAILED THE CALIBRATION TEST. IT WAS ALSO OBSERVED THAT THE ENCLOSURE OF THE SUBJECT DEVICE WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT REPORTED.