inforMED
MalfunctionCDD

ALINITY I ACTIVE B12 (HOLOTRANSCOBALAMIN) REAGENT KIT

Received May 11, 2026 · Event occurred Apr 30, 2026

Report 3002809144-2026-00121 · MDR key 25136873

Device

Generic name

Radioassay, Vitamin B12

Manufacturer

Abbott Gmbh

Catalog number

09P26-35

Lot number

12569UP00

Product problems

  • High Test Results

Patient

30 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I ACTIVE B12 RESULT GENERATED ON AN ALINITY I PROCESSING MODULE FOR ONE PATIENT SAMPLE. SID (B)(6) RESULT ON LOT 12569UP00 = 29.7 PMOL/L, RESULT ON OTHER LOT = 18.1 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. CONTINUED INFORMATION FOR SECTION E1 PHONE NUMBER: (B)(6).