MalfunctionCDD
ALINITY I ACTIVE B12 (HOLOTRANSCOBALAMIN) REAGENT KIT
Received May 11, 2026 · Event occurred Apr 30, 2026
Report 3002809144-2026-00121 · MDR key 25136873
Device
Product problems
- High Test Results
Patient
30 YR · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I ACTIVE B12 RESULT GENERATED ON AN ALINITY I PROCESSING MODULE FOR ONE PATIENT SAMPLE. SID (B)(6) RESULT ON LOT 12569UP00 = 29.7 PMOL/L, RESULT ON OTHER LOT = 18.1 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. CONTINUED INFORMATION FOR SECTION E1 PHONE NUMBER: (B)(6).