GALAFLEX LITE
Received May 11, 2026 · Event occurred Apr 21, 2026
Report 1213643-2026-00612 · MDR key 25136714
Device
Generic name
Breast - Absorbable Mesh
Manufacturer
Tepha, Inc. - 3005670760Model number
100027-33Catalog number
GFLT0022
Lot number
LXLN0001
Product problems
- Device Contamination with Chemical or Other Material
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AS REPORTED, DURING A BREAST RECONSTRUCTION PROCEDURE AFTER MASTECTOMY, A "HAIR" WAS FOUND IN THE GALAFLEX LITE PACKAGING. BASED ON THE PHOTO PROVIDED AND SAMPLE EVALUATION, FOREIGN MATERIAL (EYEBROW HAIR-LIKE) WAS OBSERVED ON TOP OF THE INNER STERILE ENVELOPE OF GALAFLEX LITE. THE HAIR MAY BE PRESENTED DURING THE MANUFACTURING PROCESS OR AT THE TIME THE STERILE PACK WAS OPENED. BASED ON THE EVENT REPORTED AND SAMPLE CONDITION THE MOST PROBABLE CAUSE OF HAIR PRESENTING IS DURING MANUFACTURING. REVIEW OF MANUFACTURING RECORDS CONFIRMS THE PRODUCT WAS MANUFACTURED TO SPECIFICATION. TO DATE, THIS IS THE ONLY REPORTED COMPLAINT FOR THIS MANUFACTURING LOT OF (B)(4) UNITS. NOTE: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
Description of Event or Problem
AS REPORTED, DURING A BREAST RECONSTRUCTION PROCEDURE AFTER MASTECTOMY, A "HAIR" WAS FOUND IN THE GALAFLEX LITE PACKAGING. ANOTHER PRODUCT WAS USED FOR THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY.