ALINITY I FREE T4 REAGENT KIT
Received May 11, 2026 · Event occurred Apr 17, 2026
Report 3005094123-2026-00221 · MDR key 25136271
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-30
Lot number
84095UD00
Product problems
- High Test Results
Patient
32 YR · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR A 32-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 9.0-19.0 PMOL/L): SAMPLE ID (B)(6) INITIAL RESULT, ON (B)(6) 2026, WAS >64.35, REPEAT RESULT WAS 19.13, REPEAT RESULTS, ON ANOTHER ANALYZER, WERE 18.29, 18.17, AND 18.39 PMOL/L. THE TSH RESULTS FOR THIS PATIENT WERE ALSO PROVIDED: 2.521, REPEAT WAS 2.631 MIU/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.