inforMED
MalfunctionGGP

HEMOSIL VON WILLEBRAND FACTOR ACTIVITY

Received Mar 20, 2012 · Event occurred Nov 15, 2011

Report 1217183-2012-00003 · MDR key 2513240

Device

Generic name

Factor Deficiency Test

Model number

NA

Catalog number

0020004700

Lot number

B11655

Product problems

  • High Test Results

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

CUSTOMER INDICATED THAT THE NUMBER OF VWF ACTIVITY TO VWF ANTIGEN RATIOS HAS BEEN INCREASING. TOO MANY RATIOS ARE GREATER THAN 1.5.

Additional Manufacturer Narrative

ATTACH IS A SYNOPSIS OF THE TESTING TO DATE AND THE INVESTIGATION THAT IS ON-GOING AT THE CONTRACT MFR, BIOKIT. A FOLLOW-UP REPORT WILL BE FILED AS SOON AS THE CONCLUSION AND RISK ASSESSMENT ARE FINALIZED. CONCLUSION: A REVIEW OF THE DIFFERENCE BETWEEN COLUMNS B AND C AS WELL AS THE RATIOS IN COLUMNS D AND E INDICATES THAT THERE ARE SOME SAMPLES FOR WHICH INCUBATION WITH BOVINE IGG CAUSED THE VWF ACTIVITY RESULT AND THUS THE RATIO TO DECREASE, WHICH SUGGESTS THE PRESENCE OF HABIA IN THESE PT SAMPLES. STATUS: BIOKIT IS CURRENTLY PERFORMING FURTHER INVESTIGATION OF THE REAGENT.