inforMED
MalfunctionMWI

CONNEX SPOT ACCY POWER MGMT STAND

Received May 8, 2026 · Event occurred Apr 13, 2026

Report 1316463-2026-00036 · MDR key 25129375

Device

Generic name

Monitor,physiological,patient(without Arrhythmia Detection Or Alarms)

Manufacturer

Welch Allyn Inc

Model number

7000-APM

Product problems

  • Sparking

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE HAS BEEN RETURNED TO BAXTER FOR INSPECTION. A FINAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem

THE CUSTOMER REPORTED THAT THE POWER CONNECTORS BECOME ERODED AT THE TIPS AFTER A PERIOD OF USE AND THIS CAN GENERATE SPARKS WHEN REMOVING THE DEVICE FROM THE WALL WHEN IT IS CHARGED. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).