MalfunctionQJZ

XLUNG KIT 230

Received May 8, 2026 · Event occurred Apr 17, 2026

Report 3012172416-2026-00036 · MDR key 25122318

Device

Generic name

Extracorporeal System For Long-term Respiratory / Cardiopulmonary Failure

Manufacturer

Xenios Ag

Catalog number

32000014

Lot number

GSXH2504

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem
  • Insufficient Device Problem Information

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.

Description of Event or Problem

A USER FACILITY REPORTED THAT DURING THE CALIBRATION OF THE P1 PRESSURE SENSING PORT, IT WAS FOUND TO BE NOT MEASURABLE. HOWEVER, THE CONSOLE INDICATED THAT ALL PRESSURE MEASUREMENT CABLES WERE OPERATIONAL AT THE SENSOR BOX. THE PRESSURE TRANSDUCERS WERE REPLACED. THE ISSUE PERSISTED AND THE P1 PRESSURE SENSING PORT REMAINED NOT MEASURABLE. THE USER EXCHANGED THE XLUNG KIT 230, ALL PRESSURES COULD BE MEASURED, AND THE SYSTEM COULD BE CALIBRATED. THERE WAS NO PATIENT INVOLVEMENT. THE COMPLAINT SAMPLE WAS AVAILABLE FOR PRODUCT EVALUATION.