inforMED
InjuryOOE

ALLY ADAPTIVE CATARACT TREATMENT SYSTEM

Received May 7, 2026 · Event occurred Apr 15, 2026

Report 3009026057-2026-00018 · MDR key 25116086

Device

Generic name

Ally Adaptive Cataract Treatment System

Manufacturer

Lensar, Inc.

Model number

N/A

Catalog number

70-00050-001

Lot number

N/A

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Unknown

  • Capsular Bag Tear

Narrative

Description of Event or Problem

ON (B)(6) 2026, DOCTOR (B)(6) REPORTED TO CAS THAT THEY EXPERIENCED A RADIAL TEAR AT THE "JUNCTION OF THE TORIC CAPSULE NUB" DURING PROCEDURE ID # (B)(6).

Additional Manufacturer Narrative

REVIEW OF PROCEDURE #(B)(6) DOES NOT REVEAL ANY EVIDENCE OF A RADIAL TEAR AT THE REPORTED LOCATION. THE CAPSULOTOMY IS OBSERVED TO BEGIN ON FRAME 3, WITH FULL BREAKTHROUGH ACHIEVED BY FRAME 7. IT IS RECOMMENDED THAT THE CAPSULAR BUTTON REMOVAL TECHNIQUE BE REVIEWED. THE SYSTEM FUNCTIONED AS DESIGNED. NO FURTHER CLINICAL FOLLOW UP IS REQUIRED.