ALINITY I FREE T4 REAGENT KIT
Received May 7, 2026 · Event occurred Apr 3, 2026
Report 3005094123-2026-00216 · MDR key 25113122
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-20
Lot number
80419UD00
Product problems
- High Test Results
Patient
72 YR · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED, NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE.
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 RESULTS FOR A 72-YEAR-OLD FEMALE PATIENT. "ON" (B)(6) 2026, ALINITY I FREE T4 RESULT FOR SID (B)(6) WAS 3.57 NG/DL, REPEATED AFTER RE-CENTRIFUGATION >5.00 NG/DL. OTHER TESTING PERFORMED: LUMIPULSE 1.12 NG/DL, COBAS 1.09 NG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.