Received Feb 8, 1993 · Event occurred Jan 23, 1993

Report 2510 · MDR key 2510

Device

Generic name

Hospital Bed

Model number

N/A

Catalog number

N/A

Lot number

N/A

Patient

90 YR

Narrative

Description of Event or Problem

RESIDENT SLIPPED BETWEEN SIDERAIL AND MATRESS. DEATH BY SUFFOCATION DUE TO BEING WEDGED BETWEEN SIDERAIL AND MATRESSINVALID DATA - REGARDING SINGLE USE LABELING OF DEVICE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.DEVICE NOT SERVICED IN ACCORDANCE WITH SERVICE SCHEDULE. NO DATA - REGARDING DATE LAST SERVICED. SERVICE PROVIDED BY: INVALID DATA. INVALID DATA - SERVICE RECORDS AVAILABILITY. INVALID DATA - REGARDING WHETHER EVENT PRESENTS IMMINENT HAZARD. DEVICE USED AS LABELED/INTENDED.DEVICE WAS EVALUATED AFTER THE EVENT. METHOD OF EVALUATION: ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED, VISUAL EXAMINATION. RESULTS OF EVALUATION: DESIGN - INADEQUATE. CONCLUSION: INTERMITTENT FAILURE DIRECTLY CAUSED EVENT, DEVICE FAILURE DIRECTLY CONTRIBUTED TO EVENT. CERTAINTY OF DEVICE AS CAUSE OF OR CONTRIBUTOR TO EVENT: YES. CORRECTIVE ACTIONS: NO DATA. INVALID DATA - ON DEVICE DESTROYED/DISPOSED OF STATUS.