HANDPIECE, ELECTRIC DERMATOME

Received May 6, 2026 · Event occurred Oct 7, 2025

Report 0001526350-2026-00547 · MDR key 25096344

Device

Generic name

Dermatome

Catalog number

00882100100

Lot number

61936628

Product problems

  • Mechanical Problem

Patient

Not reported

  • Insufficient Information

Narrative

Additional Manufacturer Narrative

THIS COMPLAINT IS RECORDED BY ZIMMER BIOMET UNDER (B)(4). G2: FOREIGN - EVENT OCCURRED IN AUSTRALIA. REVIEW OF THE MOST RECENT REPAIR RECORD DETERMINED THE UNIT WAS FOUND TO HAVE LOW MOTOR SPEED DURING TESTING. THE MOTOR AND WIRING HARNESS WERE REPLACED AND RESOLVED THE REPORTED ISSUE. REVIEW OF THE DEVICE HISTORY RECORD IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. THE ROOT CAUSE IS ATTRIBUTED TO COMPONENT FAILURE. THE EVENT WAS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem

IT WAS REPORTED THAT DURING SURGERY, THE UNIT IS NOT FUNCTIONING PROPERLY. IT IS UNKNOWN WHETHER THERE WAS ANY PATIENT IMPACT OR DELAY. IT WAS FOUND DURING THE INVESTIGATION THAT THE UNIT HAS INCONSISTENT SPEED. DUE DILIGENCE IS COMPLETE AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.