inforMED
MalfunctionGCM

VIDEO-OPTIK "ENDOEYE 3D", 30°

Received May 6, 2026 · Event occurred Apr 16, 2026

Report 9610773-2026-03573 · MDR key 25095707

Device

Generic name

Rigid Video Laparoscope

Model number

WA50082A

Product problems

  • Communication or Transmission Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD A COMMUNICATION ERROR E227. THE ISSUE WAS FOUND DURING INSTALLATION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: NO IMAGE. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: VIDEO CABLE BROKEN. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.