inforMED
InjuryMBB

SMARTSET GMV 40G US EO

Received May 5, 2026 · Event occurred Apr 21, 2026

Report 1818910-2026-08227 · MDR key 25084995

Device

Generic name

Bone Cement : Bone Cement

Catalog number

545050501

Product problems

  • Loss of or Failure to Bond

Patient

81 YR · Male

  • Insufficient Information

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT: (B)(4). INVESTIGATION SUMMARY: AN ANALYSIS OF THE PRODUCT COULD NOT BE PERFORMED SINCE A PHYSICAL SAMPLE WAS NOT RECEIVED FOR EVALUATION. AN EVALUATION OF THE MANUFACTURING RECORD COULD NOT BE PERFORMED AS THE REQUIRED PRODUCT IDENTIFICATION NUMBER WAS NOT PROVIDED TO COMPLETE THE EVALUATION. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION, THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTIVE ACTION IS REQUIRED AT THIS TIME. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. HOWEVER, IF THE PRODUCT IS RECEIVED AT A LATER DATE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. ADDED: D1, D10 (CONCOMITANT).

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT HAD ASEPTIC TIBIAL LOOSENING. TIBIAL TRAY DEBONDED AT THE CEMENT/TRAY INTERFACE. FEMUR WASN'T LOOSE. THE TIBIA AND FEMUR WERE BOTH REVISED. NO PATIENT INJURY OR DELAY. NO MORE INFO KNOWN. PRODUCT STATED TO BE LOOSE: TIBIA. LOOSENING INTERFACE: CEMENT TO IMPLANT. CEMENT MANUFACTURER: DEPUY. DOI: (B)(6) 2012, DOR: (B)(6) 2026, AFFECTED SIDE: RIGHT KNEE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. G4: "DMF# - 13704. TRADE NAME ¿ GENTAMICIN SULPHATE ACTIVE INGREDIENT(S) ¿ GENTAMICIN SULPHATE DOSAGE FORM - POWDER STRENGTH ¿ 1.0G ACTIVE IN OUR CEMENTS." IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.