VIDEO-OPTIK "ENDOEYE 3D", 30°

Received May 5, 2026 · Event occurred Apr 22, 2026

Report 9610773-2026-03510 · MDR key 25076471

Device

Generic name

Rigid Video Laparoscope

Model number

WA50082A

Product problems

  • No Display/Image

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: H2, H3, H6, H11. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. ADDITIONALLY, BROKEN CHARGED COUPLED UNIT (R-UNIT) AND INADEQUATE RESOLUTION WAS FOUND TO BE REPORTABLE DURING INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT BROKEN VIDEO CABLE LED TO THE MALFUNCTION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

NO ADDITIONAL INFORMATION FROM THE CUSTOMER.

Description of Event or Problem

IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE BECAME DARK. THE ISSUE FOUND DURING THE NEPHRECTOMY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.