AQUABEAM ROBOTIC SYSTEM
Received Apr 29, 2026 · Event occurred Apr 6, 2026
Report 3012977056-2026-00097 · MDR key 25033282
Device
Product problems
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Male
- Perforation
Narrative
Description of Event or Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY TO TREAT SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS BECAME AWARE THAT, TWO (2) DAYS POST AQUABLATION THERAPY, THE PATIENT WAS PRESENTED TO THE EMERGENCY ROOM WITH AN INTRAPERITONEAL BLADDER PERFORATION. THE BLADDER WAS REPAIRED BY ANOTHER ON-CALL SURGEON, AND THE PATIENT HAS SINCE BEEN DISCHARGED. IT WAS REPORTED THAT THE CAUSE OF THE INTRAPERITONEAL BLADDER PERFORATION IS UNKNOWN. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.
Additional Manufacturer Narrative
THE AQUABEAM ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS CURRENTLY IN THE POSSESSION OF THE USER FACILITY. THE INVESTIGATION OF THIS EVENT CONSISTED OF A REVIEW OF THE TREATMENT LOG FILES, DEVICE HISTORY RECORD (DHR), AND INSTRUCTIONS FOR USE (IFU). THE AQUABEAM ROBOTIC SYSTEM'S TREATMENT LOG FILES WERE REVIEWED. THE REVIEW OF THE TREATMENT LOG FILES REVEALED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AS NO MALFUNCTION WAS OBSERVED DURING AQUABLATION THERAPY. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR AB2000/SERIAL NUMBER (B)(6) WAS CONDUCTED, WHICH CONFIRMED THAT THERE WERE NO NON-CONFORMANCES, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. THE REVIEW INDICATED THAT THE DEVICE MET ALL DESIGN AND MANUFACTURING SPECIFICATIONS WHEN RELEASED FOR DISTRIBUTION. THE AQUABEAM ROBOTIC SYSTEM'S INSTRUCTIONS FOR USE (IFU), IFU0101-00, REV. E, WAS REVIEWED. 4.3. WARNINGS: PROCEDURE AS WITH ANY SURGICAL UROLOGIC PROCEDURE, POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE INCLUDE: BLADDER OR PROSTATE CAPSULE PERFORATION. THE AQUABEAM ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. THE AQUABEAM ROBOTIC SYSTEM'S IFU LISTS BLADDER PERFORATION AS A POTENTIAL RISK OF AQUABLATION THERAPY. IT WAS REPORTED THAT THE CAUSE OF THE INTRAPERITONEAL BLADDER PERFORATION IS UNKNOWN. THE PATIENT HAS SINCE BEEN DISCHARGED. BASED ON THE INFORMATION RECEIVED, PLUS A REVIEW OF THE TREATMENT LOG FILES, DHR, AND IFU, THE EVENT IS CONSIDERED NOT TO BE DEVICE RELATED. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.