inforMED
MalfunctionQHM

ALINITY S HTLV I/II REAGENT KIT

Received Apr 28, 2026 · Event occurred Apr 7, 2026

Report 1415939-2026-00012 · MDR key 25020764

Device

Generic name

Serology Tests Used For Infectious Disease Screening Of Blood And Plasma Donatio

Catalog number

06P07-61

Lot number

80048MM00

Product problems

  • False Positive Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER OBSERVED FALSE REPEAT REACTIVE ALINITY S HTLV I/II RESULTS FOR MULTIPLE DONOR SAMPLES THAT WERE NEGATIVE FOR THE NAT TESTING. THE FOLLOWING INFORMATION WAS PROVIDED. THE CUTOFF IS 1.00 S/CO. SID (B)(6); INITIAL RESULT = 3.44 S/CO REACTIVE, REPEAT TESTING = REACTIVE, REPEAT- REACTIVE, RETENTION TUBE = NON-REACTIVE, NAT= NEGATIVE. SID (B)(6); INITIAL RESULT = 2.10 S/CO REACTIVE, REPEAT TESTING = REACTIVE, REPEAT- REACTIVE, RETENTION TUBE = NON-REACTIVE, NAT= NEGATIVE. SID (B)(6) ; INITIAL RESULT = 3.15 S/CO REACTIVE, REPEAT TESTING = REACTIVE, REPEAT- REACTIVE, RETENTION TUBE = NON-REACTIVE, NAT= NEGATIVE. SID (B)(6); INITIAL RESULT = 2.65 S/CO REACTIVE, REPEAT TESTING = REACTIVE, REPEAT- REACTIVE, RETENTION TUBE = NON-REACTIVE, NAT= NEGATIVE. SID (B)(6) ; INITIAL RESULT = 2.38 S/CO REACTIVE, REPEAT TESTING = REACTIVE, REPEAT- REACTIVE, RETENTION TUBE = NON-REACTIVE, NAT= NEGATIVE. SID (B)(6): INITIAL RESULT = 2.71 S/CO REACTIVE, REPEAT TESTING = REACTIVE, REPEAT- REACTIVE, RETENTION TUBE = NON-REACTIVE, NAT= NEGATIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.