inforMED
MalfunctionFMZ

GIRAFFE INCUBATOR CARESTATION

Received Apr 28, 2026 · Event occurred Apr 1, 2026

Report 3036470679-2026-00050 · MDR key 25018108

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • Inaccurate Delivery

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 SERVO MODULE WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.

Description of Event or Problem

IT WAS REPORTED THAT IT WAS NOT POSSIBLE TO INCREASE THE PERCENTAGE OF OXYGEN DELIVERY. THERE WAS NO PATIENT INVOLVEMENT.