ALINITY I FREE T4 REAGENT KIT
Received Apr 27, 2026 · Event occurred Apr 15, 2026
Report 3005094123-2026-00193 · MDR key 25001027
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-20
Product problems
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
A CONFERENCE ABSTRACT BY CULLEN, M.R., AND LEE, G.R., ¿EVALUATION OF FREE THYROXINE AND HIGH SENSITIVITY CARDIAC TROPONIN DUPLICATE TESTING,¿ CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2026; 64(4): EA78. DOI: 10.1515/CCLM-2026-0148, DESCRIBES THE USE OF ROUTINE DUPLICATE TESTING AS A RISK MANAGEMENT STRATEGY TO DETECT ANALYTICAL ERRORS IN ELEVATED ALINITY I FREE T4 RESULTS GENERATED BY THE ALINITY I PROCESSING MODULES. DUPLICATE TESTING WAS PERFORMED FOR ALL ELEVATED ALINITY I FREE T4 RESULTS >20 PMOL/L. RESULTS WERE ASSESSED USING A PREDEFINED ACCEPTABILITY CRITERION OF A >20% DIFFERENCE AT CONCENTRATIONS =20 PMOL/L. NINETY-TWO DUPLICATE PAIRS WERE ANALYZED, FROM WHICH ONLY ONE DEMONSTRATED AN ANALYTICAL ERROR (1.1%). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Additional Manufacturer Narrative
SECTION E: E1: ADDRESS 1: COMPLETE ADDRESS INFORMATION: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.