BENCHMARK ULTRA SYSTEM
Received Apr 27, 2026
Report 2028492-2026-01611 · MDR key 25000064
Device
Generic name
Slide Stainer, Automated
Manufacturer
Ventana Medical Systems Inc.Model number
BENCHMARK ULTRACatalog number
05342716001
Product problems
- Computer Software Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
1. SUMMARY OF INVESTIGATION RESULTS: THE INVESTIGATION IS ONGOING. 2. SUMMARY OF FOLLOW UP/CORRECTIVE ACTIONS TAKEN: THE INVESTIGATION IS ONGOING. 3. DEVICE RETURNED TO MANUFACTURER? NO. 4. DEVICE LABELED "FOR SINGLE USE?" NO. 5. DEVICE REPROCESSED AND REUSED? NO.
Description of Event or Problem
1. DESCRIBE THE EVENT: THERE WAS AN ALLEGATION OF A SOFTWARE ISSUE WITH A BENCHMARK ULTRA SYSTEM. AFTER A SLIDE WAS CORRECTLY PROGRAMMED FOR THE MSH6 STAINING PROTOCOL, THE SOFTWARE CRASHED. THE SOFTWARE WAS RE-STARTED AND THE RUN WAS ABORTED. WHEN THE SOFTWARE DISPLAYED THE MSH6 SLIDE, IT SHOWED THAT IT WAS PROGRAMMED FOR THE "VIMENTIN" STAINING PROTOCOL INSTEAD. 2. PATIENT AGE RANGE: ASKU. 3. PATIENT WEIGHT RANGE: ASKU. 4. PATIENT SEX BREAKDOWN: ASKU. 5. PATIENT RACE BREAKDOWN: ASKU. 6. PATIENT ETHNICITY BREAKDOWN: ASKU.