inforMED
MalfunctionMDM

SCORPION NEEDLE, KNEE

Received Apr 27, 2026 · Event occurred Apr 7, 2026

Report 1220246-2026-02463 · MDR key 24999217

Device

Generic name

Manual Instr, General Surgical

Manufacturer

Arthrex, Inc.

Catalog number

AR-12990N

Lot number

15528155

Product problems

  • Break

Patient

NA · Unknown

  • Foreign Body In Patient
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Description of Event or Problem

ON 11TH APRIL 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT 2 UNITS OF AR-12990N, SCORPION¿ NEEDLE, KNEE, WERE BROKEN WHEN THE SURGEON TRIED VERY HARD TO PASS A SUTURE THROUGH THE MENISCUS ROOT. THE PATIENT'S KNEE WAS SUPER SMALL AND SUPER TIGHT. WHEN THE NEEDLE PASSED THROUGH THE MENISCUS ROOT, THE NEEDLE TIP HIT THE FEMUR BONE AND BROKE. THIS WAS DETECTED DURING A RIGHT KNEE MENISCUS ROOT REPAIR AND CENTRALISATION PROCEDURE ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION RECEIVED ON 14TH APRIL 2026: HOWEVER, THE BROKEN PIECES WERE NOT RETRIEVED FROM THE PATIENT. AS IT FELL TO THE BACK OF THE KNEE, SURGEON DARE NOT ANYHOW POKE AS THERE'S VASCULAR BUNDLES.