EVIS LUCERA ELITE XENON LIGHT SOURCE
Received Apr 24, 2026 · Event occurred Oct 9, 2025
Report 3002808148-2026-13106 · MDR key 24985776
Device
Product problems
- Gas/Air Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: DUE TO FRONT PANEL WAS CRACKED AT THE INTERFACE, HAS LIQUID IMMERSION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE XENON LIGHT SOURCE EXHIBITED WATER INVASION IN FRONT PANEL FROM CRACKS IN THE INTERFACE. THERE WAS NO PATIENT INVOLVEMENT.