MalfunctionGCT

EVIS LUCERA ELITE XENON LIGHT SOURCE

Received Apr 24, 2026 · Event occurred Oct 9, 2025

Report 3002808148-2026-13106 · MDR key 24985776

Device

Generic name

Xenon Light Source

Model number

CLV-290SL

Product problems

  • Gas/Air Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: DUE TO FRONT PANEL WAS CRACKED AT THE INTERFACE, HAS LIQUID IMMERSION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE XENON LIGHT SOURCE EXHIBITED WATER INVASION IN FRONT PANEL FROM CRACKS IN THE INTERFACE. THERE WAS NO PATIENT INVOLVEMENT.