STIMLOC
Received Apr 24, 2026 · Event occurred Apr 21, 2026
Report 2182207-2026-01085 · MDR key 24980924
Device
Generic name
Cover, Burr Hole
Manufacturer
Medtronic NeuromodulationModel number
924256Catalog number
924256
Lot number
082224324A
Product problems
- Insufficient Device Problem Information
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H3. ANALYSIS OF THE BURR HOLE DEVICE (LOT# 082224324A) FOUND THAT ONLY THE BURR HOLE COVER BASE, CENTERING TOOL, INSERTION TOOL (GREY AND BLUE), INSERTION TOOL (CLEAR), AND A DISPOSABLE SCREWDRIVER WERE RETURNED/RECEIVED. THERE WAS NO ANOMALY OBSERVED WITH THE BURR HOLE BASE. THE BURR HOLE CLIP AND BURR HOLE CAP WERE MISSING FROM THE RETURNED ITEMS; THEREFORE, NO ANALYSIS COULD BE PERFORMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT THE LOCKING CAP COULD NOT LOCK THE LEAD.