MalfunctionGCT
EVIS LUCERA ELITE XENON LIGHT SOURCE
Received Apr 24, 2026 · Event occurred Nov 27, 2025
Report 3002808148-2026-12886 · MDR key 24979948
Device
Product problems
- Poor Quality Image
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE XENON LIGHT SOURCE HAD NOISY IMAGES. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. HOWEVER, BASED ON THE RESULTS OF THE INVESTIGATION, RUST ON THE INTERFACE IS CONSIDERED THE MOST LIKELY CONTRIBUTOR TO THE OBSERVED IMAGE NOISE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.