MalfunctionLNH

ACHIEVA 1.5T DSTREAM

Received Apr 24, 2026 · Event occurred Apr 13, 2026

Report 3042177665-2026-000050 · MDR key 24975477

Device

Generic name

Full-body Mri System, Superconducting Magnet

Catalog number

782112

Product problems

  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

PHILIPS RECEIVED A REPORT THAT THE BODY COIL GASKET WAS DETACHED FROM THE FRONT CONE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS DAMAGED IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

Additional Manufacturer Narrative

THE QBC SEAL THAT HAD COME LOOSE AND CAME OFF, WAS REPLACED AND THE SYSTEM WAS RETURNED TO THE CUSTOMER FOR USE. THE FAILURE OF THE QBC SEAL TO REMAIN IN PLACE IS CONSIDERED A MALFUNCTION. REMEDIAL ACTION: A FIELD ACTION WAS INITIATED UNDER (B)(4). A FIELD SAFETY NOTIFICATION WAS SENT TO THE CUSTOMERS INFORMING THEM ABOUT THIS POTENTIAL ISSUE. THE FIELD CORRECTION HAS SINCE BEEN MADE AVAILABLE TO THE FIELD IN DEC 2024.

Remedial action

  • Modification/Adjustment