ACHIEVA 1.5T DSTREAM
Received Apr 24, 2026 · Event occurred Apr 13, 2026
Report 3042177665-2026-000050 · MDR key 24975477
Device
Generic name
Full-body Mri System, Superconducting Magnet
Manufacturer
Philips Medical Systems Nederland B.v.Model number
SMARTPATH TO DSTREAM FOR 1.5TCatalog number
782112
Product problems
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
PHILIPS RECEIVED A REPORT THAT THE BODY COIL GASKET WAS DETACHED FROM THE FRONT CONE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS DAMAGED IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.
Additional Manufacturer Narrative
THE QBC SEAL THAT HAD COME LOOSE AND CAME OFF, WAS REPLACED AND THE SYSTEM WAS RETURNED TO THE CUSTOMER FOR USE. THE FAILURE OF THE QBC SEAL TO REMAIN IN PLACE IS CONSIDERED A MALFUNCTION. REMEDIAL ACTION: A FIELD ACTION WAS INITIATED UNDER (B)(4). A FIELD SAFETY NOTIFICATION WAS SENT TO THE CUSTOMERS INFORMING THEM ABOUT THIS POTENTIAL ISSUE. THE FIELD CORRECTION HAS SINCE BEEN MADE AVAILABLE TO THE FIELD IN DEC 2024.
Remedial action
- Modification/Adjustment