inforMED
MalfunctionCAD

TEC

Received Apr 23, 2026 · Event occurred Apr 10, 2026

Report 2112667-2026-02723 · MDR key 24972869

Device

Generic name

Vaporizer

Manufacturer

Datex-ohmeda, Inc.

Model number

850 SEV

Catalog number

NA

Lot number

NA

Product problems

  • Output below Specifications

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

GE HEALTHCARE (GEHC) REPORTED A FIELD MODIFICATION FOR REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT PER 21 CFR 806 ON 27-NOV-2024. THE FDA RECALL NUMBER IS (RECALL NO. Z-0814-2025). CUSTOMERS WERE SENT A LETTER EXPLAINING THE ISSUE AND INSTRUCTIONS FOR INSPECTING FOR CORRECT OUTPUT CONCENTRATION. GEHC WILL REPLACE ALL AFFECTED UNITS. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.

Description of Event or Problem

AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION AND REMOVAL INITIATED BY GE HEALTHCARE (GEHC) ON 27-NOV-2024, (RECALL NO. Z-0814-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE WITH REDUCED CONCENTRATION DELIVERY OF ANESTHETIC AGENT. THERE WAS NO PATIENT INVOLVEMENT.

Remedial action

  • Recall