inforMED
MalfunctionFSM

ATTUNE

Received Apr 23, 2026 · Event occurred Mar 4, 2026

Report 1818910-2026-07483 · MDR key 24967371

Device

Generic name

Instrument Cases

Catalog number

254900110

Lot number

PV0165931

Product problems

  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY- ACCORDING TO THE INFORMATION RECEIVED, IT WAS REPORTED THAT THE INSTRUMENT HAD AN UNKNOWN ALLEGATION. THIS EVENT DID NOT HAPPEN IN SURGERY. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE ATTUNE REV ASSEMBLY CASE HAD SOME BRACKETS BROKEN. ADDITIONALLY, THE COATING MATERIAL WAS FOUND PEELED OFF IN SOME BRACKETS. BREAKAGE CAN BE ATTRIBUTED TO UNINTENDED USE ERROR WHILE INTERACTING WITH THE DEVICE, POSSIBLY BY PILING HEAVY OBJECTS OVER THE TRAY, OR IMPACTING WITH OTHER HEAVY INSTRUMENTS, CAUSING EXCESSIVE STRESS ON THE HOLDERS AND BREAKAGE OVER TIME. PROPERLY HANDLING AND¿ATTENTION TO THE APPROVED USE OF THE DEVICE DIMINISHES THE RISK OF FAILURE. THE OBSERVED PEELING COATING CONDITION WAS IDENTIFIED AS AN END OF LIFE INDICATOR; DAMAGE CONSISTENT WITH REPEATED USE AND SERVICING. THE LIFECYCLE REQUIREMENTS OF THE DEVICE ARE EVENT RELATED AND DEPEND ON THE USE AND INSPECTION OF THE DEVICE IN CLINICAL PRACTICE. AS THE DEVICE CAN BE DAMAGED ON THE FIRST OR 100TH USE, THE DEVICE MUST BE PROPERLY INSPECTED PRIOR TO EACH SURGICAL USE. REFER TO THE DEVICE/COUNTRY SPECIFIC IFU FOR INFORMATION RELATED TO END OF LIFE, REPROCESSING INSTRUCTIONS, AND INSPECTION PROCEDURES. A FUNCTIONAL TEST WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE ATTUNE REV ASSEMBLY CASE WOULD HAVE CONTRIBUTED TO A DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT- THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.

Description of Event or Problem

IT WAS REPORTED THAT THE INSTRUMENTS HAD AN UNKNOWN ALLEGATION. THIS WAS FOUND WHEN INSPECTING PANS THAT WERE RETURNED FROM SURGERY.