inforMED
MalfunctionGCM

VIDEO TELESCOPE "ENDOEYE 3D", 10 MM, 0°

Received Apr 23, 2026

Report 9610773-2026-03215 · MDR key 24967283

Device

Generic name

Rigid Video Laparoscope

Model number

WA50080A

Product problems

  • Poor Quality Image

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE 3D FUNCTION AND WHITE BALANCE ARE NO LONGER WORKING. THE ISSUE OCCURRED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, H2, H3, H6, H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS NOT CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: IMAGE IS NOT DISPLAYED. BASED ON THE RESULTS OF THE INVESTIGATION, AND SINCE THE DEVICE MALFUNCTION WAS NOT CONFIRMED DURING EVALUATION, THE DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. ADDITIONALLY, THE IMAGE WAS NOT DISPLAYED DUE TO A BROKEN VIDEO CABLE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem