TRURIZECLINICAL CHAIR
Received Apr 23, 2026
Report 0001831750-2026-99246 · MDR key 24966970
Device
Generic name
Chair, Positioning, Electric
Manufacturer
Stryker Medical-kalamazooCatalog number
3333000000
Product problems
- Electrical /Electronic Property Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. 2 DEVICES WERE FUNCTIONALLY/VISUALLY INSPECTED IN THE FIELD. THE DEVICES WERE REPAIRED AND RETURNED TO USE. EVALUATION RESULTS FINDINGS (COMPONENTS/RESULTS) 2 DEVICES: POWER CORD / CURRENT LEAKAGE PROBLEM. THERE WAS NO REMEDIAL ACTION TAKEN. THIS DEVICE IS NOT LABELED FOR SINGLE USE.
Description of Event or Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.