CANNULATED SCREW SYSTEM +
Received Apr 22, 2026 · Event occurred Mar 21, 2026
Report 3012966183-2026-00003 · MDR key 24959133
Device
Generic name
7.5mm Ti Cann Screw + Ft - 95mm
Manufacturer
Tyber MedicalModel number
04.355.895Lot number
21182AD93
Product problems
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE EVENT WAS INVESTIGATED WITH THE INFORMATION PROVIDED. NO LOT INFORMATION WAS PROVIDED; THEREFORE A DEVICE HISTORY RECORD REVIEW FOR LOT-RELATED NONCONFORMITIES COULD NOT BE PERFORMED. AN EVALUATION COULD NOT BE PERFORMED, AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. IMAGES WERE PROVIDED THAT CONFIRMED THE EVENT OF THE SCREW HEAD BROKEN OFF OF THE SHAFT. WITHOUT FURTHER INFORMATION OR THE PHYSICAL DEVICE FOR EVALUATION, THE ROOT CAUSE CANNOT BE ESTABLISHED. FURTHER INVESTIGATIONS WILL BE PERFORMED SHOULD THE DEVICE BE MADE AVAILABLE FOR RETURN OR ADDITIONAL INFORMATION BECOME AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT ON (B)(6) 2026, A 7.5 CSS+ SCREW BROKE WHEN INSERTING IT INTO THE BONE. VERY LITTLE RESISTANCE NEEDED TO BREAK THE HEAD OFF. SURGERY WAS DELAYED FOR 30 MINUTES. NO PATIENT CONSEQUENCES.