inforMED
MalfunctionGES

CARBON RIB-BACK BLADES SIZE 15, STERILE

Received Apr 22, 2026

Report 3002750084-2026-00040 · MDR key 24957838

Device

Generic name

Scalpel Blade

Model number

371115

Lot number

0496979

Product problems

  • Unsealed Device Packaging

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

COMPLAINANT ALLEGES "FOUND SEVERAL 15 SURGICAL BLADES WHERE THE OUTER PACKAGING IS NOT SEALED PROPERLY."

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED FOR EVALUATION AND PICTURES WERE PROVIDED. AFTER EVALUATION, IT WAS DETERMINED THAT THE ROOT CAUSE WAS A MANUFACTURING ERROR. THE ROOT CAUSE IS MACHINE RELATED. LOSS OF REGISTRATION OCCURRED DURING THE FOIL PREPARATION, RESULTING IN THE BLADE CUTTING THE POUCH IN THE WRONG LOCATION. THIS SHOULD BE CONSIDERED THE FINAL REPORT. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.