CARBON RIB-BACK BLADES SIZE 15, STERILE
Received Apr 22, 2026
Report 3002750084-2026-00040 · MDR key 24957838
Device
Generic name
Scalpel Blade
Manufacturer
Aspen Surgical Products, Inc.Model number
371115Lot number
0496979
Product problems
- Unsealed Device Packaging
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
COMPLAINANT ALLEGES "FOUND SEVERAL 15 SURGICAL BLADES WHERE THE OUTER PACKAGING IS NOT SEALED PROPERLY."
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED FOR EVALUATION AND PICTURES WERE PROVIDED. AFTER EVALUATION, IT WAS DETERMINED THAT THE ROOT CAUSE WAS A MANUFACTURING ERROR. THE ROOT CAUSE IS MACHINE RELATED. LOSS OF REGISTRATION OCCURRED DURING THE FOIL PREPARATION, RESULTING IN THE BLADE CUTTING THE POUCH IN THE WRONG LOCATION. THIS SHOULD BE CONSIDERED THE FINAL REPORT. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.