inforMED
MalfunctionOEQ

CUSTOM KIT

Received Apr 17, 2026 · Event occurred Jan 28, 2026

Report 1721504-2026-00015 · MDR key 24918212

Device

Generic name

Convenience Kit (multiple Products)

Catalog number

K09-MS3049A

Lot number

H3330865

Product problems

  • Tear, Rip or Hole in Device Packaging

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE ACCOUNT ALLEGES THAT THERE WAS A DEFECT IN THE PACKAGING RESULTING IN INCOMPLETE STERILIZATION OF THE CONTENTS. THE DEFECT WAS IDENTIFIED DURING INCOMING INSPECTION BY A DEVICE DISTRIBUTOR. NO PATIENT INTERACTION OR INJURY TO REPORT.

Additional Manufacturer Narrative

THE SUSPECT DEVICE WAS RETURNED FOR EVALUATION. THE HOLE IS THE PACKAGING WAS OBSERVED. HOWEVER, IT IS NOT POSSIBLE TO DETERMINE WITH CERTAINTY THE ORIGIN OF THE HOLE IN THE PACKAGING; IT COULD HAVE BEEN PRESENT SINCE RECEIPT OF THE MATERIAL FROM THE SUPPLIER, OR IT MAY HAVE OCCURRED DURING THE PACKAGING PROCESS, TRANSPORT, OR THE INSPECTIONS CONDUCTED BY THE END USER. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND. THE DEVICE HISTORY RECORD WAS REVIEWED, AND NO EXCEPTION DOCUMENTS WERE FOUND. NONCONFORMANCES OF THIS NATURE ARE MONITORED BY THE OPERATION TEAM. THIS COMPLAINT WILL BE USED TO TREND FOR SIMILAR COMPLAINTS.