MalfunctionGXN

ACCUSHAPE PEEK

Received Apr 17, 2026 · Event occurred Apr 1, 2026

Report 3009196021-2026-00009 · MDR key 24913032

Device

Generic name

Peek Cranial Implant

Manufacturer

Medcad

Model number

ZB-SPK30

Catalog number

ZB-SPK30

Lot number

263023HUS

Product problems

  • Failure to Clean Adequately

Patient

77 YR · Male

  • Failure of Implant

Narrative

Additional Manufacturer Narrative

MEDCAD EVALUATED THE DEVICE AND DETERMINED THAT THE OBSERVED SHEEN ORIGINATED FROM THE MACHINING PROCESS. ADDITIONALLY, THE FEW DRAINAGE HOLES WERE FOUND TO CONTAIN PEEK MATERIAL. SURGEON WAS ABLE TO RE-SCHEDULE THE SURGERY, AND A NEW IMPLANT ALONG WITH A DUPLICATE IMPLANT, WAS PROVIDED BEFORE THE RESCHEDULED SURGERY DATE UPON CONFORMANCE.

Description of Event or Problem

PEEK IMPLANT HAD SHEEN ON IT AND DRAINAGE HOLES WERE NOT CLEANED ALL THE WAY THROUGH.

Remedial action

  • Replace