Device
Generic name
Bone Cement : Bone Cement
Manufacturer
Depuy Orthopaedics Inc UsCatalog number
545050501
Lot number
7489925
Product problems
- Loss of or Failure to Bond
Patient
74 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO ASEPTIC TIBIAL LOOSENING AT CEMENT TO IMPLANT INTERFACE. TIBIAL IMPLANT WAS POSITIONED WITH A SLIGHT ANTERIOR SLOPE. TIBIA DEBONDED AT THE CEMENT/IMPLANT INTERFACE. ALL COMPONENTS WERE REMOVED. FEMUR WAS NOT LOOSE. NO DELAYS OR PT INJURY. DOI: (B)(6) 2014. DOR: (B)(6) 2026. AFFECTED SIDE: LEFT KNEE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. "JRN COMBO PRODUCTS: DMF# (B)(4). TRADE NAME ¿ GENTAMICIN SULPHATE. ACTIVE INGREDIENT(S) ¿ GENTAMICIN SULPHATE. DOSAGE FORM - POWDER. STRENGTH ¿ 1.0G ACTIVE IN OUR CEMENTS."
Additional Manufacturer Narrative
PRODUCT COMPLAINT: (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A RECORDS EVALUATION (MRE) WAS NOT PERFORM. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. ADDED: D1, D10 CONCOMITANT.