MAGNETOM AERA
Received Apr 17, 2026 · Event occurred Apr 10, 2026
Report 3002808157-2026-00012 · MDR key 24910458
Device
Generic name
System, Nuclear Magnetic Resonance Imaging
Manufacturer
Siemens Healthineers AgModel number
10432914Product problems
- Device Handling Problem
Patient
NA · Unknown
- Limb Fracture
Narrative
Description of Event or Problem
SIEMENS HEALTHINEERS WAS NOTIFIED OF A SERIOUS INJURY ASSOCIATED WITH A MAGNETOM AERA SYSTEM. IT WAS COMMUNICATED THAT A FERROMAGNETIC IV POLE WAS ATTRACTED TO THE MAGNET. THE SYSTEM OPERATOR WAS INJURED DURING THIS EVENT, SUFFERING LOSS OF CONSCIOUSNESS FOR APPROXIMATELY 40 SECONDS, A FRACTURED FIBULA, AND A LACERATION BEHIND THE RIGHT EAR. THE OPERATOR WENT TO THE EMERGENCY DEPARTMENT AND RECEIVED SUTURES BEHIND THE EAR AND AN EXAMINATION OF THE LEG INJURY. THE OPERATOR WAS DISCHARGED WITH CRUTCHES AND A PROTECTIVE SHOE, AND WAS INSTRUCTED TO FOLLOW WITH ORTHOPEDICS IN ONE WEEK. DURING THE ORTHOPEDIC APPOINTMENT, IT WAS DETERMINED THAT SURGERY IS REQUIRED TO TREAT THE FIBULA FRACTURE.
Additional Manufacturer Narrative
H3, H6: THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON RECEIPT OF ADDITIONAL REPORTABLE INFORMATION AND/OR INVESTIGATION COMPLETION.